Infusion Pump Maintenance: Terumo TE-Series Guide

8 min read
89 views
Last updated: November 9, 2025
Infusion Pump Maintenance: Terumo TE-Series Guide

Detailed maintenance guide for Terumo TE-Series infusion pumps covering daily checks, calibration procedures, and component service.

Infusion Pump Maintenance: Terumo TE-Series Guide

The Terumo TE-Series infusion pumps require systematic preventive maintenance to ensure accurate medication delivery and patient safety. This guide provides comprehensive maintenance procedures for biomedical equipment technicians.

Overview

TE-Series pumps utilize advanced peristaltic pump mechanisms with electronic flow control. Regular maintenance prevents accuracy drift, mechanical wear, and electronic failures that could compromise patient care.

Maintenance Schedule

Daily Maintenance Tasks

Pre-Use Inspection

Essential safety check before clinical use:

Visual Inspection Points:

- [ ] External housing integrity

- [ ] Display screen clarity and function

- [ ] All buttons and controls responsive

- [ ] Power cord condition

- [ ] Battery charge indicator status

- [ ] Pump door operation

- [ ] Air sensor window cleanliness

Basic Function Test:

1. Power On Sequence:

- Connect to AC power

- Press power button

- Observe self-test completion

- Check for error messages

- Verify display brightness

2. Battery Test:

- Disconnect AC power

- Verify battery operation

- Check battery life indicator

- Test low battery alarm

- Reconnect AC power

Cleaning and Sanitization

Infection control and equipment preservation:

External Cleaning Protocol:

1. Power down pump completely

2. Clean with hospital-approved disinfectant:

- Use soft, lint-free cloths

- Clean all external surfaces

- Pay attention to high-touch areas

- Clean display screen gently

- Avoid liquid ingress into openings

3. Pump Mechanism Cleaning:

- Open pump door

- Remove any residual tubing fragments

- Clean pump head with alcohol wipe

- Inspect for wear or damage

- Close door securely

Weekly Flow Accuracy Testing

Gravimetric Method Verification

Gold standard for pump accuracy testing:

Required Equipment:

- Analytical balance (±0.1g accuracy)

- Graduated cylinder (100mL)

- Stopwatch

- Approved IV tubing sets

- Test fluid (normal saline)

Testing Procedure:

Step-by-Step Process:

1. Setup:

- Install fresh tubing set

- Prime tubing completely

- Position collection container on scale

- Zero scale with empty container

2. Flow Rate Testing:

- Program pump to test flow rate

- Start timer and pump simultaneously

- Monitor for occlusion alarms

- Stop pump at end of test period

- Weigh collected fluid immediately

3. Accuracy Calculation:

- Accuracy % = (Delivered Volume / Expected Volume) × 100

- Record results in maintenance log

- Investigate if accuracy outside tolerance

- Schedule calibration if needed

Volume Delivery Verification

Test different volume settings:

Volume Accuracy Test:

- Program pump for 50mL delivery

- Use 25 mL/hr flow rate

- Measure actual delivered volume

- Should be within ±2% of programmed volume

- Test stop function accuracy

Monthly Mechanical Inspection

Pump Head Assembly Service

Critical component requiring regular attention:

Disassembly Procedure:

1. Preparation:

- Power down pump

- Remove from service

- Gather required tools

- Document configuration

2. Pump Head Removal:

- Open pump door completely

- Remove pump head retaining screws

- Carefully lift out pump head assembly

- Note rotor position and alignment

Component Inspection:

Maintenance TypeFrequencyDurationComplexity
Basic Function CheckDaily5 minutesBasic
Cleaning & SanitizationDaily15 minutesBasic
Flow Accuracy VerificationWeekly30 minutesIntermediate
Mechanical InspectionMonthly45 minutesAdvanced
Electronic CalibrationQuarterly90 minutesExpert
Complete OverhaulSemi-annually180 minutesExpert
Test Flow RateTest DurationExpected VolumeTolerance
1 mL/hr60 minutes1.0 mL±5%
5 mL/hr60 minutes5.0 mL±5%
25 mL/hr30 minutes12.5 mL±3%
100 mL/hr15 minutes25.0 mL±3%
ComponentInspection PointsAction Required
Pump RotorsWear, cracks, debrisClean, replace if worn
Pressure PlateAlignment, wear marksAdjust, replace if damaged
SpringsTension, corrosionReplace if weak/corroded
BearingsSmooth operationLubricate, replace if rough

Rotor Replacement Procedure:

1. Remove old rotors using proper tool

2. Clean rotor mounting surfaces

3. Install new rotors with correct orientation

4. Apply specified torque (2.5 Nm)

5. Verify smooth rotation

6. Reassemble pump head

Motor System Evaluation

Electric motor and drive system:

Motor Performance Testing:

- Measure motor current draw (normal: <150mA)

- Check motor speed regulation

- Test encoder feedback accuracy

- Verify motor mounting security

- Listen for unusual noises

Drive System Inspection:

- Check gear train for wear

- Verify proper lubrication

- Test backlash and play

- Inspect coupling integrity

- Document any abnormalities

Quarterly Electronic Calibration

Flow Sensor Calibration

Ensures accurate flow detection:

Calibration Equipment:

- Certified flow reference standard

- Oscilloscope for signal analysis

- Precision voltage source

- Manufacturer calibration software

Sensor Calibration Procedure:

1. Signal Analysis:

- Monitor sensor output waveform

- Check signal amplitude and frequency

- Verify sensor supply voltage (5V ±0.25V)

- Test signal-to-noise ratio

2. Calibration Adjustment:

- Access service calibration mode

- Run pump at known flow rate

- Adjust calibration factor to match reference

- Test multiple flow rates for linearity

- Save calibration parameters

Pressure Monitoring System

Critical for occlusion detection:

Pressure Sensor Testing:

1. Static Pressure Test:

- Apply known pressure (0-300 mmHg)

- Verify sensor output accuracy

- Check pressure alarm thresholds

- Test pressure release function

2. Dynamic Response Test:

- Create pressure ramps

- Measure sensor response time

- Should be <1 second to 90% reading

- Test over full pressure range

Display and Interface Calibration

User interface accuracy:

Touchscreen Calibration (if applicable):

- Access calibration mode

- Follow screen touch sequence

- Verify touch accuracy at all points

- Test screen sensitivity

- Calibrate display brightness

Button Response Testing:

- Test all buttons for proper response

- Check tactile feedback

- Verify button legends legibility

- Test button lighting (if present)

Semi-Annual Complete Overhaul

Major Component Service

Comprehensive system restoration:

Power Supply Service:

1. AC Power Section:

- Measure input voltage regulation

- Test power factor correction

- Check for electromagnetic interference

- Verify grounding integrity

2. DC Power Rails:

- Measure all DC voltages (±5V, ±12V, 24V)

- Check ripple and noise levels

- Test load regulation

- Verify over-current protection

3. Battery System:

- Test battery capacity (should be >80% rated)

- Check charging circuit operation

- Verify backup power duration (minimum 4 hours)

- Replace battery if capacity degraded

Software and Firmware Updates

Keep system current with latest versions:

Update Procedure:

1. Backup Current Configuration:

- Save all calibration parameters

- Document current software versions

- Export pump settings and logs

2. Software Installation:

- Download latest firmware from manufacturer

- Follow installation procedure exactly

- Verify update completion

- Test all functions after update

3. Configuration Restoration:

- Restore calibration parameters

- Verify all settings correct

- Test pump operation thoroughly

- Update documentation

Safety Precautions

⚠️ CRITICAL SAFETY REQUIREMENTS

- Never perform maintenance with power applied

- Use only manufacturer-approved parts and lubricants

- Follow lockout/tagout procedures during service

- Test all safety systems after maintenance

- Verify pump accuracy before returning to service

- Maintain sterile technique when handling pump components

- Never defeat or bypass safety interlocks

- Ensure proper grounding during electrical testing

Performance Documentation

Maintenance Records

Complete documentation required:

Daily Inspection Logs:

- Date, time, and technician ID

- Inspection checklist completion

- Any anomalies noted

- Actions taken

- Next service due

Calibration Records:

- Calibration date and procedure

- Reference standards used

- Before/after measurements

- Adjustments made

- Acceptance criteria met

- Technician certification

Component Service History:

- Parts replaced and dates

- Service procedures performed

- Performance trends

- Failure analysis reports

- Cost tracking

Troubleshooting Common Issues

Flow Rate Inaccuracy

Most common pump problem:

Diagnostic Steps:

1. Check tubing installation

2. Verify tubing type compatibility

3. Inspect pump head for wear

4. Test flow sensor calibration

5. Check for air bubbles

6. Verify proper priming

Correction Methods:

- Recalibrate flow sensor

- Replace worn pump components

- Use correct tubing sets

- Improve priming technique

Alarm Malfunctions

Safety system integrity issues:

Common Alarm Problems:

- False occlusion alarms

- Intermittent air detection

- Battery alarm errors

- Flow rate deviation alerts

Resolution Approach:

- Clean optical sensors

- Adjust alarm thresholds

- Replace faulty sensors

- Update alarm software

External References

1. Terumo TE-Series Service Manual (terumo-infusion.com/service-support)

2. FDA Infusion Pump Safety Communications (fda.gov/medical-devices/infusion-pumps)

3. OSHA Healthcare Equipment Maintenance Standards (osha.gov/healthcare)

*Last Updated: November 9, 2025*

*Reading Time: 10 minutes*

---

⚠️ Important: This guide summarizes safe repair practices based on industry best practices and manufacturer guidelines. Always consult the official manufacturer manual for model-specific procedures. Medical equipment repair should only be performed by qualified personnel.

Share this article

Was this helpful?

Important Safety Notice

This guide summarizes safe repair practices based on industry best practices and manufacturer guidelines. Always consult the official manufacturer manual for model-specific procedures. Medical equipment repair should only be performed by qualified personnel.

Related Articles