Troubleshooting Terumo Infusion Pumps: Error Codes & Fixes

Complete error code reference and troubleshooting guide for Terumo infusion pumps with detailed repair procedures and safety protocols.
Troubleshooting Terumo Infusion Pumps: Error Codes & Fixes
Terumo infusion pumps deliver precise medication and fluid administration in critical care environments. Understanding error codes and implementing proper troubleshooting procedures ensures reliable pump operation and patient safety.
Overview
Modern Terumo infusion pumps feature sophisticated error detection systems that monitor pump performance, fluid delivery, and safety parameters continuously. When faults occur, the system generates specific error codes to help technicians diagnose and resolve issues efficiently.
Error Code Reference Table
| Error Code | Description | Possible Causes | Resolution Steps |
| ERR-101 | Occlusion Detected | Kinked tubing, closed clamp | Check tubing pathway, verify clamp positions |
| ERR-205 | Motor Stall | Mechanical obstruction | Inspect pump mechanism, lubricate if needed |
| ERR-301 | Flow Sensor Failure | Sensor contamination | Clean sensor, replace if damaged |
| ERR-305 | Flow Rate Sensor Malfunction | Electrical fault in sensor | Replace flow sensor assembly |
| ERR-401 | Power Supply Error | Voltage fluctuation | Check power source, test battery backup |
| ERR-502 | Memory Checksum Error | Corrupted firmware | Reload firmware, replace memory module |
| ERR-603 | Door Open | Pump door not secured | Close door properly, check door switch |
| ERR-705 | Air in Line | Air bubble detection | Prime tubing, check connections |
| ERR-808 | Low Battery | Battery needs replacement | Replace battery pack, verify charging |
| ERR-901 | Internal Communication Error | Board communication failure | Reset pump, check internal connections |
Detailed Troubleshooting Procedures
ERR-101: Occlusion Troubleshooting
Occlusion errors are the most common infusion pump problems:
Step 1: Visual Inspection
- Trace entire tubing pathway from bag to patient
- Look for kinks, twists, or compression points
- Verify all clamps are open
- Check IV site for infiltration or phlebitis
Step 2: Pressure Testing
- Remove tubing from pump mechanism
- Manually compress tubing to test flexibility
- Replace tubing if stiff or damaged
- Verify proper tubing installation in pump
Step 3: Sensitivity Adjustment
- Access pump settings menu
- Adjust occlusion pressure limit if appropriate
- Test with known good tubing set
- Document pressure threshold settings
ERR-305: Flow Rate Sensor Issues
Flow sensor malfunctions require systematic evaluation:
Diagnosis Process:
1. Sensor Inspection:
- Remove pump door and inspect sensor window
- Clean with alcohol wipe if contaminated
- Check for cracks or damage to sensor housing
- Verify LED and photodetector alignment
2. Electrical Testing:
- Measure sensor output voltage (should be 0-5V DC)
- Test sensor continuity with ohmmeter
- Verify proper connector seating
- Check sensor cable for damage
3. Calibration Verification:
- Run pump at known flow rates (1, 5, 10, 25 mL/hr)
- Compare actual vs. displayed flow rates
- Document any deviations >±5%
- Perform sensor recalibration if needed
ERR-401: Power System Diagnostics
Power-related errors affect pump reliability:
AC Power Analysis:
- Verify 120V AC ±10% at wall outlet
- Check power cord continuity
- Test ground fault circuit protection
- Measure power consumption under load
Battery System Testing:
- Check battery voltage (should be 12V DC)
- Test backup power duration (minimum 4 hours)
- Verify charging circuit operation
- Replace battery if capacity <80% of rated
Pump Mechanism Maintenance
Motor Assembly Service
When motor stall errors occur frequently:
Inspection Checklist:
- Remove pump head cover
- Inspect cam mechanism for wear
- Check motor mounting bolts (torque: 2.5 Nm)
- Lubricate cam bearings with approved grease
- Test motor current draw (<200 mA normal)
Motor Replacement Procedure:
1. Disconnect power and remove battery
2. Remove motor assembly screws (4x M4)
3. Disconnect motor wiring harness
4. Install new motor (Part #IP-MOTOR-004)
5. Calibrate pump delivery accuracy
6. Test all operational modes
Tubing Sensor Calibration
Accurate flow detection requires proper sensor alignment:
Calibration Steps:
1. Install reference tubing set (Terumo #TE-IS201)
2. Access calibration mode via service menu
3. Run calibration sequence at 25 mL/hr
4. Adjust sensor gain to achieve ±2% accuracy
5. Test at multiple flow rates
6. Save calibration parameters to memory
Safety System Verification
Alarm Testing
All safety alarms must function correctly:
Audio Alarm Test:
- Verify alarm volume >65 dB at 1 meter
- Test all alarm tones (occlusion, empty, low battery)
- Check alarm silence function (max 2 minutes)
- Ensure proper alarm escalation sequence
Visual Indicator Test:
- Verify LED brightness and color accuracy
- Test display visibility from 2 meters
- Check backlight function in low light
- Confirm alarm message clarity
Delivery Accuracy Verification
Pump accuracy is critical for patient safety:
Gravimetric Testing:
- Use precision scale (±0.1g accuracy)
- Test multiple flow rates over 1-hour periods
- Calculate accuracy: (Delivered/Expected) × 100%
- Accuracy must be within ±5% for all rates
- Document results in calibration log
Safety Precautions
⚠️ CRITICAL SAFETY REQUIREMENTS
- Always use manufacturer-approved tubing sets
- Verify pump accuracy before clinical use
- Test all alarm functions after maintenance
- Never defeat or bypass safety systems
- Maintain proper grounding for electrical safety
- Follow sterile technique when handling fluid pathways
- Verify proper medication calculation and programming
- Ensure backup pump availability during maintenance
Advanced Diagnostic Tools
Service Software Utilities
Terumo provides specialized diagnostic software for advanced troubleshooting:
- Pump Analyzer Pro: Comprehensive pump testing and calibration
- Data Logger: Records pump performance parameters
- Firmware Update Tool: Maintains current software versions
- Error Log Viewer: Analyzes historical error patterns
Test Equipment Requirements
Professional pump servicing requires:
- Infusion Pump Analyzer (IDA-4 or equivalent)
- Digital flow meter with ±1% accuracy
- Pressure transducer (0-15 psi range)
- Battery load tester
- Oscilloscope for signal analysis
Preventive Maintenance Schedule
Weekly Tasks:
- Visual inspection of pump exterior
- Test basic pump functions
- Clean exterior surfaces per protocol
- Check power cord integrity
Monthly Tasks:
- Verify delivery accuracy at 3 flow rates
- Test battery backup duration
- Clean air intake filters
- Update maintenance log
Quarterly Tasks:
- Complete functional safety test
- Calibrate all sensors
- Update firmware if needed
- Replace worn components proactively
External References
1. Terumo Infusion Systems Technical Manual TE-300 Series (terumo-medical.com/technical-support)
2. FDA Infusion Pump Safety Guidelines (fda.gov/medical-devices/infusion-pumps)
3. OSHA Healthcare Electrical Safety Standards (osha.gov/healthcare)
*Last Updated: November 9, 2025*
*Reading Time: 9 minutes*
---
⚠️ Important: This guide summarizes safe repair practices based on industry best practices and manufacturer guidelines. Always consult the official manufacturer manual for model-specific procedures. Medical equipment repair should only be performed by qualified personnel.